Vee Tac Presents.....
Hemodynamics is the study of the dynamic behavior of blood. As blood flows from chamber to chamber,
as valves open and close, and as the myocardium contracts and relaxes, pressures are generated in various
parts of the heart. These cardiovascular pressures can be measured and monitered through catheters whose tips are placed in the atria, pulmonary artery or systemic arteries.
These are called "hemodynamic lines".
Hemodynamic lines have several uses. They enable you to sample venous and arterial blood without having to stick a patient over and over. They provide a way to monitor various waveforms, which can provide clues
to patient status. The combination of pulmonary, arterial, and systemic arterial lines can be used to calculate cardiac output.
Most important, these lines enable you to monitor directly various cardiac pressures. Interpretation of these pressures can guide you and the physician in planning and evaluating
therapy in shock, fluid overload or deficit, cardiac failure, and other conditions.
PRESSURE MEASUREMENTS
The most important cardiac pressure is that of the left ventricle., because it is a major determinant of systemic perfusion. The pressure in the left ventricle just before systole is called the
left ventricular end-diastolic pressure or LVEDP. This pressure reflects the compliance of the left ventricle - it's ability to receive blood from the left atrium during diastole. When the left
ventricular compliance decreases, the LVEDP rises. MI and left ventricular failure are two examples of when left ventricular compliance decreases.
CORRELATION OF PRESSURES
There is a close correlation between LVEDP and other cardiac pressures. In the presence of a normal mitral valve, LVEDP is reflected by left atrial pressure or LAP. In the person with a normal mitral valve and normal lung function, the LVEDP is also reflected by the pressure in the pulmonary capillary
bed,pulmonary capillary wedge pressure or PCWP) and the pressure in the pulmonary artery at the end of diastole. This latter pressure is sometimes referred to as the pulmonary artery end diastolic pressure or PAEDP.
Remember, this concept only hold true for patients with a normal mitral valve and no pulmonary disease.
Left arterial pressure can be monitored at bedside, but a LAP line can be dangerous because it provides a direct path for air or clots to enter the left ventricle and become systemic emboli. The pulmonary capillary and pulmonary arterial pressures can be monitored at the bedside with a balloon-tipped
catheter placed in the pulmonary artery. With the balloon deflated, one can measure pulmonary artery systolic, diastolic, and mean pressures with the catheter. When the balloon is inflated, it wedges the catheter in a small distal branch
of the pulmonary artery. The pressure recorded is that reflected back from the left atrium through the pulmonary capillary bed. This pressure is the
Pulmonary capillary wedge pressure or PCWP.
VALUES FOR NORMAL RESTING CARDIAC PRESSURES
Note..these can vary somewhat from institution to institution..
-right atrium mean 0-8 mm Hg; A wave: 2-10 mm Hg; V wave: 2-10 mm Hg
-right ventricle systolic 15-30 mm Hg; end diastolic: 0-8 mmHg
-pulmonary artery systolic 15-30 mm Hg; end diastolic: 3-12 mm Hg
-wedge A wave 3-15 mm Hg; V wave: 3-12 mm Hg; mean: 5-12 mm Hg
-AVO2 difference (mL/L) 30-50
-cardiac output (L/minute) 4.0-6.5 (varies with patient size)
-cardiac index (L/minute/m2) 2.6-4.6
-pulmonary vascular resistance (dynes - second - cm-2) 20-130
-systemic vascular resistance (dynes - second - cm-2) 700-1600
CRITICAL VALUES
Pressure tracings may be virtually diagnostic of certain conditions.
-Mitral stenosis is associated with a pressure gradient in
diastole across the mitral valve (wedge or left atrial pressure vs left
ventricular pressure). A large V wave in the pulmonary artery wedge
tracing may be seen with mitral regurgitation, since the amplitude of
the V wave is affected by left atrial filling from the pulmonary veins as
well as the regurgitant volume from the left ventricle.
-Decreases in right atrial
pressure, pulmonary capillary wedge pressure, and cardiac
index/output can indicate hypovolemia.
-In cases of elevated right atrial
pressures with low wedge pressures and low cardiac
index/output (especially in the face of an inferior wall myocardial
infarction) one may suspect right ventricular involvement and failure.
-Pulmonary congestion due to left ventricular failure or volume
overload will increase the pulmonary artery wedge pressure (ie,
congestion usually occurs at a wedge pressure in excess of 18 mm Hg
and frank pulmonary edema occurs with a wedge pressure in the
upper twenties and above).
-Cardiogenic shock and pulmonary edema
are characterized by signs of hypoperfusion, with hemodynamic data
including systemic hypotension, markedly decreased cardiac index
less than 2.1 L/minute/m2, and elevated wedge pressures, often well above 18
mm Hg.
-Septic shock is also characterized by clinical signs of
hypoperfusion, but may be differentiated from cardiogenic shock by
certain hemodynamic data which often include a normal or near
normal wedge pressure, an elevated cardiac index/output, and a
marked decrease in systemic vascular resistance.
Caution should be
exercised in that these parameters are only general guidelines, and
during the course of a patient's illness, such information may not
always be exact. As always, history and physical examination are
critical in the diagnostic assessment of each patient. The catheter can
aid with diagnostic dilemmas, but is most useful as a management tool.
Pulmonary artery catheters can also be useful in the diagnosis of
ventricular septal defects by sampling O2 saturations as the catheter is
advanced from the great veins to the right atrium to the right ventricle
and out into the pulmonary artery. An oxygen "step-up" from the right
atrium to the right ventricle of approximately 10% is indicative of left
to right shunting.
In the appropriate setting of acute myocardial
infarction and sudden deterioration after a stable course, this diagnosis
may be a consideration; right heart catheterization is one method to
establish the diagnosis. Other causes of an O2 step-up include
coronary fistula draining into the RV, primum atrial septal defects, and
pulmonic insufficiency with a patent ductus arteriosus.
Cardiac
tamponade is another diagnosis which can be documented by
pulmonary artery catheter measurements. Rising intrapericardial
pressures interfere with diastolic filling of the heart. Marked increases
in the end-diastolic pulmonary artery (PA), right ventricular (RV) ,
and right atrial (RA) pressures to the same value ("equalization of the
pressures") strongly suggest tamponade. Somewhat similar findings
may be seen with constrictive and restrictive diseases.
Pulmonary hypertension and increased
pulmonary vascular resistance can suggest such diagnoses as
pulmonary embolism or even mitral stenosis. Care must be taken in
the interpretation of all hemodynamic data derived from the catheter.
INTERPRETING PRESSURE DATA
PLEASE REMEMBER...
1. Compare the values obtained to the patient's normal values rather than an arbitrary standard. If the patient has undergone cardiac Cath within the past few months,
pressures obtained at that time may be used as baselines. If not, you must predict general values on the basis of your knowledge of the so-called normal values and your patient's pathology.
For example, you would expect the patient with a narrowed tricuspid valve to have an elevated CVP. The patient with COPD probably would have both high CVP and PA pressures.
2. Single readings are not as important as the trend of values.
3. Consider the pressures in relation to each other. If one pressure is measured with a manometer and another with a transducer, you may want to convert them to the same scale.
To convert millimeters of mercury (Hg) to centimeters of water, multiply by 1.36. Remember that abnormal values are not always due to a primary pathology of the monitored chamber. For example, and elevated CVP in association with normal or low PA pressures suggests that the
cause lies between these two sites, that is, with the pulmonary valve, right ventricle, or tricuspid valve.
4. Remember that a normal valve does not necessarily indicate an absence of pathology. For example, a patient may have a normal CVP but be intensely vaso-constricted due to hypovolemia.
Step up to the blackboard please, and meet our patient. He has just had an MI less than one week ago, and now
it appears he is going into heart failure. He is in serious need of our monitoring his hemodynamic status constantly and closely.
In ordere to do this, we are using a transducer, which is an instrument that converts
pressure waves into electrical energy so they can be displayed on an oscilloscope. We are doing this because his pressure has been too high
for the water manometer. His flush system consists of heparinized 5% Dextrose and he is obtaining this by a countiuous
low-flow flush device. This is most desirable because it is a closed system...and while it has a continuous low flow, if a rapid flow is needed,
you can pull on the "tail" of the device and flush the system without breaking sterility!
PULMONARY ARTERIAL (PA) LINES
Aortic Pressure
Pumping by the heart results in the development of pressure in the aorta and the arteries. If
pressure in the aorta is recorded over time a pressure wave can be observed:
Many factors influence the aortic pressure waveform. Consider the following example:
Greater ventricular filling (more filling time) resulted in greater
systolic pressure.
Can you think of a basic law of the heart which this situation reflects?
How about the Frank/Starling mechanism. Starling stated that "the energy
of contraction is a
function of the length of the muscle fibre." So the greater the filling of
the ventricles the stronger the
subsequent systolic contraction.
Other factors such as aortic valve condition, compliance (elasticity) of
the aorta (related to age
and disease), vascular resistance, cardiac output and technical
considerations of recording can affect
the arterial pressure waveform.
Pulmonary Artery Pressure
The pulmonary artery pressure waveform is similar in form to, but
generally of lesser magnitude
than, the aortic pressure waveform.
The type of catheter that allows you to monitor these waveforms is generally referred to by the name
of one specific brand of catheter, the Swan-Ganz pulmonary artery catheter.
Insertion of the Swan-Ganz Catheter
In general, Swan-Ganz catheterization is indicated when
measurement of right atrial, pulmonary artery, and pulmonary artery
occlusive pressures will significantly alter patient management. The
threshold for performing this procedure varies considerably amongst
clinicians; some authorities feel this technique is overutilized and is
indicated in only rare circumstances.
Contraindications
-severe, uncorrectable coagulopathy
-presence of a left bundle branch block (LBBB) on EKG;
-placement of a right heart catheter may lead to complete heart
block (A-V dissociation) if an underlying LBBB is present
-local infection at the skin insertion site
-severe hypothermia; in this situation the myocardium is highly
irritable and prone to malignant arrhythmias induced by the
catheter
-inadequate monitoring equipment; continuous EKG monitoring
with blood pressure measurements is necessary during catheter
insertion
-patient refusal
Patient Preparation
Technique and risks of the procedure are explained to the patient.
When patient is comatose or disoriented, the appropriate guardians
should be contacted. Catheterization may be safely performed in an
intensive care unit, specialized procedure room with telemetry and
fluoroscopy, or a formal Cardiac Catheterization Laboratory. A
standard emergency room or regular nursing floor is generally not
equipped for this procedure. No specific patient preparation is
required and often this procedure is performed on an urgent basis.
Whenever possible, aspirin and nonsteroidal anti-inflammatory agents
should be discontinued in advance, but this is not absolutely
necessary. Effects of heparin or warfarin, however, should be
reversed prior to catheterization. If an underlying coagulopathy is
suspected (eg, disseminated intravascular coagulation,
thrombocytopenia), appropriate laboratory studies should be
obtained immediately including a platelet count and PT/PTT. In most
cases parenteral sedation is unnecessary; however the use of agents
such as meperidine (Demerol) is at the physician's discretion.
Complications
-balloon rupture
-conduction disturbance (ie, new right bundle branch block 5%)
-arrhythmias (3% ventricular tachycardia, 2% ventricular
fibrillation)
-pulmonary infarction/pulmonary hemorrhage
perforation or rupture of the pulmonary artery
-knotting of the catheter
-thrombosis of a blood vessel (ie, 1% to 2% superior vena cava
syndrome)
-pulmonary emboli
-infection (0% to 5%)
-blood loss, including hemothorax, retroperitoneal bleed, etc
-inadvertent arterial puncture (6% femoral)
-pneumothorax and tension pneumothorax (0% to 6%)
-valvular trauma
-disconnection of the introducer apparatus with disappearance
into the vein
Equipment
-I.V. pole and pressure monitor manifold, pressure monitor
-normal saline (250-500 mL) with heparin (1000 units) for flush
-pressure bag
-pressure tubing
-stopcocks (3-way)
-cutdown tray (for peripheral approach)
-vein introducer kit
-Swan-Ganz catheter kit
-1% lidocaine for local anesthesia
-bowl of sterile saline (flush and balloon integrity check)
-suture
-instrument set
-3 and 5 mL syringes
-25-gauge needle for anesthesia
-gloves, gowns, masks
-sterile dressing kit (surgical drapes)
-bedside table on which to place instruments
-telemetry monitor for heart rate and rhythm automatic blood
-pressure cuff, A-line
-Betadine scrub
Technique
Swan-Ganz catheterization can take place via a variety of approaches
including internal jugular vein, subclavian vein, femoral vein, or
brachial vein. The last of these approaches most commonly entails
direct visualization of the brachial vein from a cut-down exposure.
The procedure should be performed in a closely monitored setting,
enabling constant recording of heart rate, rhythm, and frequent blood
pressure readings, usually an intensive care unit. The procedure may
be performed at the patient's bedside with or without the assistance of
fluoroscopic guidance. Sterile technique is required for catheter
insertion. The skin at the site of approach is most commonly prepped
with a Betadine scrub. Often, if the internal jugular or subclavian
veins are utilized, the patient is placed in a Trendelenburg position to
assist with central venous distension and ease of access. The
physician should scrub and wear gown, mask, and gloves. The patient
is then draped with sterile sheets (most institutions drape the patient
from head to toe, while others require a sterile field only at the site of
access). The patient should be cooperative for catheter insertion. If a
patient is uncooperative or becomes uncooperative during the
procedure, sedation may be given at the discretion of the physician.
Upon initiation of the procedure, the skin and subcutaneous tissue is
infiltrated with lidocaine (1%) and a small gauge needle. Deeper
tissues may then be infiltrated with lidocaine for the comfort of the
patient. A thin gauged needle (21-gauge, 112") is usually attached to
a 5 mL syringe and used to localize the vessel of interest for a central
venous approach. Once the vessel has been located, a large gauge
needle (16- or 18-gauge) is then attached to a syringe and placed into
the vessel following the course of the "finder needle." When blood is
aspirated easily into the syringe, the syringe is disconnected from the
needle and a flexible guidewire is threaded through the needle into the
vein. Wire placement can cause a variety of complications, most often
ventricular ectopy. If an increase in ectopy is observed, the guidewire
should be withdrawn several centimeters. Once the guidewire has
been passed into the vessel, the needle is removed from the patient.
At no time should the physician lose control of the tip of the
guidewire. Failure to control the guidewire can cause serious
complications and death if lost in the patient. Once the needle is
removed, a dilator is advanced over the guidewire and through the
skin, to facilitate passage of a venous introducer. The introducer
should be flushed with heparinized saline prior to insertion to avoid air
emboli. Once the tract along the guidewire is dilated, the dilator
should be slipped off the guidewire (maintaining guidewire position in
the vein). The introducer and dilator can then be put together as a unit
(dilator within introducer) and slid over the guidewire into the vein,
again taking care to control the tip of the guidewire outside the
patient's body. After the placement of the introducer and guidewire
assembly, the guidewire and dilator should be removed from the
patient. This leaves only the venous introducer sheath within the
patient. At this point, if the introducer has a side port lumen, venous
blood should be aspirated and the introducer then flushed. If blood
cannot be aspirated via a side-port lumen, the introducer is incorrectly
placed and must be reinserted. No blood should come from the
center of the introducer since this piece is usually accompanied by a
one-way ball valve which does not allow blood leakage. The
introducer should then be secured to the patient's skin with sutures.
When the venous introducer has been placed, the Swan-Ganz
catheter can then be inserted. Prior to catheter insertion, the balloon
tip should be checked under sterile water for leaks and the catheter
flushed. The catheter should then be connected to the appropriate
pressure monitoring lines and flushed again via the pressure tubing to
ensure that the catheter is bubble-free and that a column of
uninterrupted fluid exists from the tubing through the tip of the
catheter. The catheter can then be guided via the introducer, through
the central venous system, through the right atrium, right ventricle,
pulmonary artery, and into the wedge position. The catheter usually
passes smoothly through the circulation, with the aid of the inflated
balloon at its tip. The catheter should never be withdrawn with the
balloon inflated. Catheter position can be ascertained by pressure
wave forms, although fluoroscopy can be quite helpful in guiding the
catheter into the wedge position. A chest radiograph is usually
obtained after catheter insertion to verify position, as well as to rule
out the possibility of pneumothorax if the subclavian or internal jugular
approach was utilized.
ARTERIAL OR A-LINES
Arterial lines are catheters placed in systemic arteries to facilitate recording of continuous
accurate data about blood pressure in a patient who is hemodynamically unstable, and to allow frequent sampling of arterial blood gases without the need for
repeated arterial sticks. Arterial lines are commonly placed percutaneously in the radial, brachial, or femoral arteries.
The normal arterial waveform has a sharp upstroke and a more gradual downstroke
with an evident DICROTIC NOTCH, due to a small rise in pressure that occurs at the time of aortic valve closure.
End diastole should be seen very clearly...
Insertion of The A-Line
Procedure Commonly Includes
Insertion of an indwelling catheter directly into the arterial circulation
for continuous blood pressure (BP) monitoring.
Indications
May be divided into three categories:
-hemodynamic monitoring of the unstable patient (acutely
hypotensive or hypertensive) including those on vasopressor or
vasodilator agents
-multiple sampling of arterial blood, particularly in the
mechanically ventilated patient
-determination of cardiac output (less common)
Contraindications
Poor collateral circulation around the artery to be cannulated
constitutes a relative contraindication. Thrombus formation at the
catheter site is common and can result in distal extremity ischemia if
collaterals are inadequate. Also, coagulopathies, systemic
anticoagulation (eg, heparin), and interventional thrombolysis are
considered contraindications and reversal may be required.
Patient Preparation
The risks and benefits of the procedure are explained. After the site of
cannulation is selected by the physician, the area is prepared using
povidone-iodine scrub for a minimum of 30 seconds. A sterile
technique should be maintained.
Aftercare
Meticulous care is required to avoid line-related infections.
Recommendations by the Centers for Disease Control include:
-handwashing prior to any manipulation of the system
-applying topical antiseptics to the insertion site immediately
after catheter is placed
-covering the site with sterile dressing
-recording date of catheter insertion and each dressing change
-daily inspection of catheter site
-replacing sterile dressing every 48-72 hours with new antibiotic
ointment
-flushing of line using normal saline in a closed flush system
-changing flush solution every 24 hours
-changing arterial line site every 4 days or less
-removing catheter promptly at the first sign of infection
Complications
Estimates of significant complications range from 15% to 40%.
Thrombosis is the most frequent complication. Incidence of
thrombosis increases if:
-the catheter is left in place more that 3-4 days
-a large diameter catheter is used
-multiple puncture attempts are required
-hypotension, decreased cardiac output, atherosclerosis, or
hypothermia are present
-prolonged pressure is required to control bleeding after
catheter removal; thrombosis rate under optimal conditions is
approximately 5% to 8%; symptomatic occlusion requiring
surgery is much less <1%)
Infectious complications are also frequent, with the catheter serving
as either a primary or secondary site of bacteremia. Factors
predisposing to infection include prolonged (more than 4 days)
catheter insertion, the use of cutdown for insertion, local inflammation,
and infection from a secondary source. Other complications include
hemorrhage or hematoma formation, pseudoaneurysms, vasovagal
reactions, and local skin necrosis. Distal embolization of small clots or
air may occur if improper line-flush technique is used.
Equiptment
Varies somewhat depending on artery selected. A 19- or 20-gauge
teflon catheter-over-needle is used in most instances. 16 cm catheters
are used for femoral and axillary sites, shorter (114" to 2") catheters
are used for radial, brachial, and dorsalis pedis sites. If the Seldinger
technique is used, a flexible guidewire is also needed. Other
equipment includes sterile gloves, hair covers, povidone-iodine, 1%
lidocaine without epinephrine, and 3-0 or 4-0 silk suture and suture
equipment.
Technique
The radial artery is generally considered the site of choice; alternate
sites include femoral, axillary, brachial and dorsalis pedis arteries. For
radial artery cannulation, the presence of collateral flow must first be
established using the modified Allen test. Following this, the wrist is
dorsiflexed 60 degrees and using a sterile technique 1% lidocaine is used to
infiltrate overlying skin. Catheter-over-needle is inserted at a 30 degree
angle to skin and advanced until arterial blood is seen in the needle
hub. The needle is held fixed while the surrounding catheter is
advanced into the artery. The needle is removed and the catheter hub
is attached to the connecting tubing. After suturing the catheter in
place, a wrist board may be used to stabilize the neutral wrist
position. The Seldinger technique may be used for larger arteries.
Here, the artery is located with a simple 20-gauge needle. Once
arterial blood is returned, a flexible guidewire is passed through the
needle; the needle is removed and the teflon catheter is threaded over
the guidewire into the artery.
Data Acquired
Graphic waveform of arterial pressure, with pressure on the vertical
axis (mm Hg) and time on the horizontal axis
Normal Arterial Pressure Tracing
The peak of each waveform represents the systolic blood pressure
and the trough represents the diastolic blood pressure (in mm Hg).
Normal values for blood pressure obtained by arterial cannulation are
slightly higher than those obtained by routine sphygmomanometry,
ranging from 5-20 mm Hg higher. This is due to a combination of
physiologic and technical factors, reviewed elsewhere. If indirect
pressure readings (ie, cuff pressures) are greater than arterial line
readings, instrument error is likely. The entire system (tubing,
calibration, seals, catheter, etc) should be carefully inspected; the
transmitted arterial waveforms may also appear "damped," further
suggesting technical error. A normal "square wave" response is also
shown in Figure A. This waveform is seen whenever the tubing system
is flushed. Most monitoring systems are equipped with a "flush valve"
which can be opened and closed rapidly (routinely performed by
nursing staff). A rapid-velocity stream flows through the tubing,
removing bubbles and debris. The resulting waveform is by nature
artifactual, but abnormalities in its configuration suggest underlying
technical problems.
Damped Arterial Pressure Tracing
In normal individuals, peak systolic blood pressures vary somewhat
with respiration, a finding difficult to appreciate with bedside
sphygmomanometry, but easily observed with direct arterial blood
pressure monitoring. When a healthy person inspires, there is a
transient fall in blood pressure. On the blood pressure monitoring
screen, this appears as a "dip" in the pressure tracings, which returns
to baseline during expiration. The maximum drop in systolic blood
pressure (pulsus paradoxus) should not exceed 8-10 mm Hg. Values
less than this are physiologic and should not be confused with cardiac
tamponade.
CRITICAL Values
Cutoff values for hypertension, as defined in textbooks, are the same
for blood pressure obtained by arterial cannulation and routine
sphygmomanometry. A "hypertensive urgency" is characterized by
marked elevations in diastolic (and sometimes systolic) blood
pressure, accompanied by retinal hemorrhages, exudates, and
papilledema. End-organ damage is likely within several days if blood
pressure is not adequately controlled. In a "hypertensive emergency"
(malignant hypertension), the retinal findings described are present
along with such alarming features as acute renal failure, seizures,
blurred vision, mental status deterioration, stroke, and congestive
heart failure. End-organ damage is already apparent. Although both
hypertensive urgencies and emergencies show marked blood pressure
elevations (eg, diastolic blood pressure greater than 120-140 mm Hg), there are
no precise cutoff values. These syndromes should not be arbitrarily
diagnosed or excluded on the basis of arterial line blood pressure
readings alone; they are complex clinical diagnoses. Similarly, no
black-and-white cutoff values exist for defining hypotension. Most
physicians would consider a systolic blood pressure in the 70-80 mm
Hg range abnormal if the individual was previously healthy. However,
systolic blood pressures in the 80-90 mm Hg range are not unheard
of in the patient with end stage cardiac disease or on multiple
vasodilatory agents. Conversely, a "normal" systolic blood pressure of
110 mm Hg may indicate significant hypotension in the dialysis patient
whose baseline is 200 mm Hg. A drop in systolic blood pressure
during inspiration >10 mm Hg is significant. This increased
paradoxical pulse may be seen in cardiac tamponade, severe asthma,
pulmonary embolism, and other conditions. Arterial cannulation is
useful in monitoring the patient with cardiac tamponade, but is seldom
used to make the diagnosis. Disparity in blood pressure readings
between direct and indirect measurements greater than 20 mm Hg may occur in
shock states. This is due to reflex peripheral vasoconstriction
(increased systemic vascular resistance). Korotkoff sounds may be
barely audible when direct measurement of central arterial pressures
are low-normal. Large discrepancies may also be seen in patients with
severe peripheral atherosclerosis (arteriosclerosis obliterans), where
systolic pressure drops off dramatically distal to a luminal occlusion. It
should be emphasized that inaccuracies may occur in both direct and
indirect systems. Clinical importance should be placed on the trends in
blood pressure values, regardless of the system used.
Limitations
Accuracy is limited by errors introduced by the equipment, which
transforms mechanical energy (pulse) into electrical energy (tracing).
Factors such as the natural frequency of the transducer, clamping, and
compliance may cause artifact. Other sources of error include
improper leveling of equipment, improper assembly, and air in the
tubing.
Additional Information
Arterial cannulation is generally considered a procedure of low
technical difficulty. The true difficulty lies in avoidance of thrombosis
and infection and careful patient selection.
THE CVP LINE
The large veins (superior and inferior vena cavae) run into the right
atrium. A catheter inserted into
the jugular vein and passed down towards the right atrium, and connected
to a pressure transducer
measures the central venous pressure, which is essentially identical
to the right atrial pressure since
there are no valves between the right atrium and the large veins.
While most nurses, from med/surg to critical care, have assisted doctors in the insertion
of CVP lines, it is important to know the type of doctor you are going to assist with this procedure
in order to know how to prepare. Please check out and study the following link for this information, and then
return to this page by hitting your BACK key...
VERY IMPORTANT LINK
PLEASE CLICK HERE
COMMON PROBLEMS
Arterial lines and PA lines both share some common problems that all nurses must be aware of, these being
damping, spurius readings thrombosis, and infection. Exsanguination also can occur if a stopcock is accidently left open
after an arterial blood sample is drawn. The blood in the artery is under such high pressure that a patient can lose a significant amount of blood,
even if the connection just comes loose. For this reason, limbs with arterial lines are ALWAYS uncovered and pressure alarms should be set to alert you
if accidental disconnection occurs. In addition, when the line is removed, you should maintain firm pressure on the site for at least 5 minutes to prevent hematoma formation
due to high intravascular pressure.